Recall Z-1243-2007 ยท LLZ System, Image Processing, Radiological

Cerner Corp

Cerner Millennium RadNet Auto Launch Study and Auto Launch Report software functionalities.

DateSep 25, 2007 (initiated Aug 23, 2007)
StatusTerminated
ReasonDefects in the Auto Launch functionality make it possible for a mismatch of patient data.
Root causeSoftware design
ActionThe recalling firm initiated recall via telephone on 8/23/07 to their customers who had the 2007 version in their production environment, explained the reason for recall, and informed them they should turn off the affected functionality. The firm posted details of the situation on their website in two different documents (referred to as a Priority Review Flash), one dated 8/29/07 and the other dated 8/31/07. The 8/31/07 document was also mailed to their customers via certified mail on 9/4/07. The corrected software (EP28822 ver.1) was made available to the firm's clients on 9/7/07 via their
QuantityDownloaded by 9 customers nationwide.
DistributionDistribution was made to health care facilities located in LA, MO, NJ, MS, MI, NE, ND, and AR.