Recall Z-0580-2008 ยท MMI Immunoassay Method, Troponin Subunit

Siemens Medical Solutions Diagnostics

ADVIA Centaur TnI-Ultra Assay, Troponin Immunoassay, 500 tests - Part/Catalog # 02790309, and 100 tests - Part/Catalog # 02789602. For in vitro diagnostic use, Siemens Medical Solutions Diagnostics, Tarrytown, NY 10591

DateJun 5, 2008 (initiated Jun 7, 2007)
StatusTerminated
ReasonIncorrect results: Falsely elevated Troponin values have been reported which are inconsistent with the patient's clinical picture and test negative by other Troponin assays.
Root causeMaterial/Component Contamination
ActionTwo Support Bulletins (078N0361-01 & 078N0361-02) were sent by e-mail to all affected Siemens Diagnostics Branches to inform them of this issue. The first one was sent on 6/07/07 and the second one was sent on 7/05/07. Two Customer Bulletins (078D0647-01 & 078D0651-01) were sent by e-mail to the Siemens Diagnostics Branches for communication with the affected customers. The first one was sent on 6/07/07 and the second one was sent on 7/05/07. These bulletins state that if a false positive is suspected, the sample should be diluted and retested.
QuantityDomestic: 31,497 kits; Foreign: 27,321 kits
DistributionWorldwide Distribution - USA, Argentina, Austria, Australia, Brazil, Canada, Chile, Colombia, Egypt, France, Germany, Greece, Hong Kong, India, Israel, Italy, Malaysia, Morocco, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Swede
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K053020TNL-ULTRA ASSAY FOR THE ADVIA CENTAUR SYSTEM