Recall Z-1237-2007 ยท MRG System, Test, Vitamin D

Diasorin Inc.

25-Hydroxyvitamin D 125 I RIA Kit, in vitro diagnostic. REF:68100E DiaSorin Stillwater, Minnesota 55082-0285, U.S.A.

DateSep 20, 2007 (initiated Aug 21, 2007)
StatusTerminated
Reason25 Hydroxyvitamin D RIA kits may contain the incorrect lot of tracer.
Root causeOther
ActionConsignees were notified by phone on 8/21/07 followed by a written notification by Fax or email. The letter described the product and problem and recommended to inspect the kits and to destroy all tracer vials of lot number 726023. They also requested the consignee to return the findings by fax or mail.
Quantity117 kits
DistributionWorldwide, including USA, CANADA, BRAZIL, INDIA, QATAR, SWEDEN, URUGUAY, ARGENTINA, and JORDAN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953567INCSTAR 25-HYDROXYVITMIN D 1251 RIA
Incstar Corp., Sep 25, 1996