Recall Z-0113-2008 ยท IYO System, Imaging, Pulsed Echo, Ultrasonic

Ge Healthcare

GE Healthcare Voluson E8 ultrasound scanner with Software version 6.2.0 and 6.2.1 and DICOM stations with these software versions; GE Healthcare 9900 West Innovation Drive, Wauwatosa, WI 53226, USA.

DateNov 7, 2007 (initiated Jul 6, 2007)
StatusTerminated
ReasonWhen a spectrum of a Pulsed Wave Doppler shows an angle correction unequal to 0 is reloaded from the archive or the image is sent to a DICOM station, this angle correction is mistakenly disregarded resulting in an underestimation of the flow velocities.
Root causeSoftware design
ActionGE Healthcare sent customer letters on July 15, 2007 to affected customers. The letter included a short term mitigation recommendation and a long-term correction. The long-term correction will be software upgrade provided free of charge to all affected customers.
Quantity521 Units
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061682GE VOLUSON E8 ULTRASOUND SYSTEM
General Electric Co., Jul 12, 2006