ClearanceBrief
Recall Z-0060-2008 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Philips Medical Systems North America Co. Phillips

Philips Medical System - Endura X-Ray Systems (Release 2) with Remote Control Option

DateNov 10, 2007 (initiated Aug 23, 2007)
StatusTerminated
ReasonSystem Shut Down -- Unexpected Shut Down during use of the Optional Remote Control.
Root causeSoftware design
ActionOn 8/23/07, the firm sent out a Safety Notification letter, via certified mail. On 8/29/07, the firm sent out an URGENT - DEVICE CORRECTION letter via regular mail, dated 8/20/07. Both letters advised customers that if the device shuts down during use, it can be restarted within about 1 minute by switching the system off and then on again, using the system on/off buttons. The letter(s) states that software is being developed to correct the problem.
Quantity37 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.