Recall Z-0133-2008 ยท KQC Stains, Hematology

EMD Chemicals Inc

Harleco May and Grunwald's Stain; 0.17% w/v; Methanol Solution UN1230; EMD Chemicals Inc. An Affilifate of Merck KGaA Damstadt, Germany; 480 South Democrat Road; Gibbstown, NJ 08027; 1 liter and 4 liter poly bottles

DateNov 3, 2007 (initiated Jul 30, 2007)
StatusTerminated
ReasonOut of Spec -- The Batch (Lot) was incorrectly released after being rejected for out of specification of high azure and eosin absorbance at the end of the run and particulate matter at the end of the run.
Root causeEmployee error
ActionEMD Chemicals via fax and mailed an Important Product Recall letter dated 7/30/07. The letter notified end use consignee that after conducting an investigation it has determined that Harleco May and Grunwald's Stain; 0.17% w/v Item #660/75 % 660/85 has unacceptable levels of azure and eosin absorbance and undissolved dye matter. Users were requested to discontinue shipping and return product. Wholesalers were advised this is a Customer Level Product Recall and requested to sub recall product sold after 7/5/07.
Quantity26 units (1 liter), 1 unit (4 liter)
DistributionProduct was released to 2 distributors; one in the US, one in Canada