Recall Z-0459-2008 ยท JJF Analyzer, Chemistry, Micro, For Clinical Use

Roche Diagnostics Corp.

Roche Diagnostics LightCycler Capillaries (100uL) for use in conjunction with the LightCycler analyzers; Catalog #: 03337090001, plus any capillaries received at the time of installation of the LightCycler instruments or systems **, Roche Diagnostics Corporation, Indianapolis 46250.(**LightCycler 2.0 System, Catalog Number: 04391039001; Batch Number Range: 1415471-1415933; LightCycler 2.0 Instrum

DateDec 12, 2007 (initiated Aug 30, 2007)
StatusTerminated
ReasonFalse Negative Results: Air bubbles in the lens or the area surrounding the lens of the LightCycler 100 uL capillaries may cause false negative results. The affected capillaries showed up to a 40% reduction in signal
Root causeProcess design
ActionConsignees were notified by Urgent Medical Device Recall letter dated 8/30/07, to discontinue use of the 100uL capillaries, plus all of those received at installation, and to disposed of them per local guidelines.
Quantity119 boxes of 96 capillaries each
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033734LIGHTCYCLER INSTRUMENT VERSION 1.2
Roche Diagnostics Corp., Dec 17, 2003