ClearanceBrief
Recall Z-1232-2007 ยท CHI Lipase-Esterase, Enzymatic, Photometric, Lipase

Abbott Laboratories Inc.

Clinical Chemistry Lipase, List Numbers: 7D80-20 and 7D80-30

DateSep 18, 2007 (initiated Jun 1, 2007)
StatusTerminated
ReasonLipase results are falsely elevated when a MULTIGENT Acetaminophen test is pipetted immediately before a Lipase test from the same sample cup or tube on the ARCHITECT c8000.
Root causeOther
ActionThe firm initiated the recall on 06/01/2007 by sending recall letters with attached Customer Reply form via FedEx to each direct account that received the recalled product. Customers were informed of the issue and given instructions for modification to the SmartWash feature in the Lipase Assay Configuration in order to prevent Acetaminophen from contaminating Lipase.
Quantity7260
DistributionWorldwide, including USA, Australia, Brazil, Canada, Chile, El Salvador, Dominican Republic, Ecuador, Germany, Hong Kong, Japan, Mexico, New Zealand, Panama, Puerto Rico, Singapore, South Korea, Thailand, and Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.