Recall Z-1229-2007 ยท MPB Catheter, Hemodialysis, Non-Implanted

C.R. Bard Access Systems, Inc

Guidewire, 70cm x 1mm, REF 5583705, Sterile, Single Use, BARD Access Systems, Hemodialysis Catheter, Salt Lake City, UT 84116.

DateSep 20, 2007 (initiated Aug 10, 2007)
StatusTerminated
ReasonMetal guidewires used in Hemodialysis Catheters may have been reverse loaded within the dispenser. Potential exists for the stiff end of the guidewire to be inserted into the patient instead of the flexible end.
Root causeOther
ActionConsignees were notified by letter on 08/10/2007 and asked to remove and return all unused affected product to Bard.
Quantity7225 units
DistributionWorldwide, including USA, Canada, Latin America, and Europe (Austria, Belgium, Finland, France, Germany, Great Britain, Holland, Italy, Norway, Spain, and Sweden).