Recall Z-0007-2008 ยท JJX Single (Specified) Analyte Controls (Assayed And Unassayed)

Tosoh Bioscience Inc

4 glass vials per box. 2/4ml vials of Sample Diluting Solution & 2/2ml vials of Calibration Verification Material. BetaHCG Calibration Verification Test Set, in vitro diagnostic.

DateOct 3, 2007 (initiated Jun 22, 2007)
StatusTerminated
ReasonFirm discovered that a deterioration value of approximately 20% was found when the product of this lot was measured. Tosoh AIA confirmed that deterioration of the product occurred over time and resulted in a deviation from the assigned value.
Root causeNonconforming Material/Component
ActionLetter, dated July 2, 2006. Immediately examine inventory and quarantine product. If product was further distributed, the direct accounts are to notify their customers including a copy of the original recall letter and Notification of Recall response card.
Quantity47 Boxes
DistributionNationwide.