Recall Z-0321-2008 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corp

Terumo Advanced Perfusion System, 1 Roller Pump, 4 inch diameter; Catalog No. 801040, Terumo Cardiovascular Systems, Ann Arbor, MI 48103

DateDec 1, 2007 (initiated Sep 10, 2007)
StatusTerminated
ReasonFailure to operate; The pump may fail to operate, and be difficult to hand crank, due to an internal electrical short condition.
Root causeProcess control
ActionConsignees were notified via an Urgent Medical Device Correction letter dated 9/10/07, which provided them guidance in the event the problem occurs and advised them they would be contacted to schedule replacement of the circuit boards.
Quantity16
DistributionWorldwide Distribution-USA including states of Alabama, Arizona, Florida, Illinois, Indiana, Mississippi, Nebraska, North Carolina, Pennsylvania, and Virginia, and countries of Belgium, Hong Kong and United Arab Emirates.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022947ADVANCED PERFUSION SYSTEM 1 (100/120 VOLT),ADVANCED PERFUSION SYSTEM 1 (220/240 VOLT, ROLLER PUMPS (6, 4 INCH
Terumo Cardiovascular Systems Corp., Mar 20, 2003