Recall Z-0260-2008 ยท DQY Catheter, Percutaneous

Boston Scientific Target

Boston Scientific 6F Guider Softip XF Guiding Catheter, 6 French, 40 degree angle, 100cm usable length, Material Number (UPN) M003101420, Catalog Number: 10142, Single-use Disposable Guide Catheters, Made in USA: 5905 Nathan Lane, Plymouth, MN 55442 USA

DateNov 24, 2007 (initiated Jul 11, 2007)
StatusTerminated
ReasonMislabeling: Catheters were found to be mislabeled (100 cm labeled devices actually contains 90cm length catheter)
Root causeLabeling mix-ups
ActionFirm notified consignees by Urgent Medical Device Recall Letter dated 11 Jul 07 requesting return of product.
Quantity49 units
DistributionWorldwide Distribution-USA, EU, Japan, South Korea, Mexico, Colombia, and Hungary.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K980453SCHNEIDER GUIDER SOFTIP GUIDING CATHETERS
Boston Scientific Scimed, Inc., Aug 11, 1998