Recall Z-0132-2008 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

VNUS Medical Technologies Inc

VNUS ClosureFAST catheters, Model Number: CL7-7-100, VNUS Medical Technologies, Inc., San Jose, CA 95138

DateDec 15, 2007 (initiated Mar 6, 2007)
StatusTerminated
ReasonSterility (package integrity) Compromised: Product sterility may be compromised due to a pinhole under the pouch foldover.
Root causePackage design/selection
ActionThe recalling firm has issued customer notification letters by mail on 03/06/2007. The firm is requesting that consignees discontinue their use of the device and return recalled product . The firm will provide replacements.
Quantity92 units
DistributionNationwide: Product was distributed to 18 health care providers and medical centers throughout the US in the following states: TX, UT, CA, CT, MI, NY, MI, OR, WA, VA, PA, and NJ.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061373VNUS CLOSUREFAST CATHETER, MODELS CF7-7-60 AND CF7-7-100
Vnus Medical Technologies, Inc., Aug 1, 2006