Recall Z-1155-2007 ยท IYE Accelerator, Linear, Medical
Northwest Medical Physics Equipment Inc
ISOLOC Software Version 4.2, Part Number NT-NW-425-520. Image guided radiation treatment software.
| Date | Aug 9, 2007 (initiated Nov 30, 2004) |
| Status | Terminated |
| Reason | When ISOLOC Version 4.2 unit is set to centimeter and the user clicks the Reenter button in the digitizer window, the coordinates of the markers and target in the anatomic imaging system erroneously displays values ten times greater than the correct values. |
| Root cause | Software design |
| Action | Software to fix the described problem was written and customers contacted by e-mail on 11/30/2004, that a software patch, titled 11302004, was available to their customers on the firm's internet website. |
| Quantity | 83 |
| Distribution | Worldwide, including USA, Korea, Spain, and UK. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K904908 | STEREOTACTIC IRRADIATION SYSTEM Northwest Medical Physics Center, Jun 6, 1991 |