Recall Z-1154-2007 ยท IYE Accelerator, Linear, Medical

Northwest Medical Physics Equipment Inc

ISOLOC Software Version 6.0, Part Number NT-NW-425-520. Localization software for radiotherapy treatment system.

DateAug 8, 2007 (initiated Dec 30, 2004)
StatusTerminated
ReasonWhen ISOLOC version 6.0 software was used in the Anatomic Landmark mode, the couch moves display in the Couch Translator window was incorrect.
Root causeSoftware design
ActionSoftware to fix the described problem was written and customers contacted by e-mail on 12/30/2004 that a software patch, titled 12302004 was available to their customers on the firm's internet website.
Quantity79
DistributionWorldwide, including USA, Korea, Spain, and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K904908STEREOTACTIC IRRADIATION SYSTEM
Northwest Medical Physics Center, Jun 6, 1991