Recall Z-1153-2007 ยท IYE Accelerator, Linear, Medical
Northwest Medical Physics Equipment Inc
ISOLOC Software Version 6.0, Localization software for determining treatment machine position (to position patients) for high accuracy radiotherapy.
| Date | Aug 8, 2007 (initiated Apr 11, 2005) |
| Status | Terminated |
| Reason | When ISOLOC version 6.0 was used in the centimeter mode, the display in the couch moves window had a rounding error relative to the Portal Image Marker window of up to 0.5 mm. |
| Root cause | Software design |
| Action | Software to fix the described problem was written and customers contacted by e-mail on 04/11/2005, that a software patch, titled 04112005, was available to their customers on the firm's internet website. |
| Quantity | 84 |
| Distribution | Worldwide, including USA, Korea, Spain, and UK. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K904908 | STEREOTACTIC IRRADIATION SYSTEM Northwest Medical Physics Center, Jun 6, 1991 |