Recall Z-0989-2007 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

GE OEC Medical Systems, Inc

GEHC OEC 9900 ESP-15 Elite C-Arm fluoroscopic X-ray system, GE Healthcare, Salt Lake City, UT.

DateJul 12, 2007 (initiated Jul 6, 2007)
StatusTerminated
ReasonThe Air Kerma Rate displayed value and cumulative air kerma displayed value could exceed the allowable error under certain imaging conditions.
Root causeOther
ActionConsignees were notified by certified letter on 07/06/2007. They were told that a service rep would contact them to provide a label and an addendum to their operator's manual. A permanent software solution is under development and will be provided when it becomes available.
Quantity22 units
DistributionWorldwide, including USA, France, Singapore, Sweden, and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041931OEC OLYMPUS MOBILE FLUOROSCOPY SYSTEM WITH INTEGRATED NAVIGATION
Ge Oec Medical Systems, Inc., Aug 26, 2004