Recall Z-0130-2008 ยท GEX Powered Laser Surgical Instrument

AMS Innovative Center - San Jose

GreenLight HPS System, Surgical Laser System and Accessories. Part Numbers 0010-0070 and 0010-0075 (international units), AMS Innovative Center, San Jose, CA 95134.

DateMar 20, 2008 (initiated Apr 17, 2007)
StatusTerminated
ReasonComponent missing: Some products were issued without thermal protection switches.
Root causeProcess design
ActionAMS issued Important: Medical Device Correction letters, dated April 17, 2007, to its US and international consignees by mail, instructing them to return an acknowledgement form. The letter informs customers that some HPS systems may lack a motor overload switch, that the firm believes that there is no immediate safety issue, and installation of the MOL switch will be conducted during normal preventive maintenance. AMS expects to complete installation of the switches by end of August 2007.
Quantity169 units
DistributionWorldwide Distribution - USA, Italy, France, Germany, Spain, UK, Denmark, Sweden, Australia, India, Japan, Greece, Switzerland, Saudi Arabia, Thailand, Turkey, Singapore, Taiwan/Hong Kong (China), and Austria.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010284LYRA G SERIES SURGICAL LASER SYSTEM (SL SERIES Q-SWITCHED ND:YAG CONFIGURATION)
Laserscope, May 1, 2001
K062719GREENLIGHT HPS SERIES SURGICAL LASER SYSTEM & ACCESSORIES
Laserscope, Dec 1, 2006