Recall Z-1190-2007 ยท IYO System, Imaging, Pulsed Echo, Ultrasonic

Volcano Corporation

Volcano s5 Imaging system, Ultrasound pulsed echo imaging system. Product consists of the CPU, control console, monitor, printer, and patient interface module (PIM)

DateAug 23, 2007 (initiated Apr 2, 2007)
StatusTerminated
ReasonDuring emissions testing, it was discovered that the product exceeds emissions limits for electromagnetic compatibility.
Root causeOther
ActionThe firm notified its customers by letter beginning April 2, 2007, and started replacing the monitor with appropriate replacement parts.
Quantity351 units worldwide
DistributionProduct was distributed to health care facilities nationwide and to distributors in Europe (via subsidiary), Japan (via subsidiary), Latin America and Asia (third party distributors).
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051920VOLCANO S5 IMAGING SYSTEM, MODEL 804200-001