Recall Z-1199-2007 ยท GKZ Counter, Differential Cell

Abbott Laboratories

Abbott CELL-DYN 3500 Hematology Analyzer

DateAug 29, 2007 (initiated May 11, 2007)
StatusTerminated
ReasonThe product failed electrostatic discharge testing, which could result in product failure.
Root causeOther
ActionMandatory Technical Service Bulletins (TSB) are being issued to Field Service to upgrade all instruments in the field. All customers will be contacted by Field Service to schedule this mandatory upgrade. The upgrades are expected to take 12 months.
Quantity3,429 units, all models
DistributionWorldwide, including USA, Puerto Rico, Canada, Mexico, Argentina, Australia, Bahamas, Brazil, Chile, China, Colombia, Costa Rica, Dominican Republic, El Salvador, Ecuador, Germany, Guatemala, Hong Kong, Jamaica, Japan, New Zealand, Panama, Paraguay, Peru, Singapore, South Korea, Taiwan, Thailand, Ur
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K913305CELL-DYN 3000A MULTI-PARAMETER AUTO HEMA ANALZYER
Unipath , Ltd., Nov 22, 1991
K955715CELL DYN 3500 MULTI-PARAMETER AUTOMATED HEMATOLOGY ANALYZER (MODIFICATION)
Abbott Laboratories, Oct 8, 1996
K960427ABBOTT CELL-DYN 3500R SYSTEM
Sequoia Turner Corp., May 2, 1996
K980614ABBOTT CELL-DYN 3000 SERIES HEMATOLOGY SYSTEMS
Abbott Laboratories, Apr 24, 1998
K991605CELL-DYN 3500 SYSTEM WITH IMMATURE RETICULOCYTE FRACTION (IRF), CELL-DYN 3700 SYSTEM WITH IMMATURE RETICULOCYTE FRACTION
Abbott Laboratories, Jul 9, 1999