ClearanceBrief
Recall Z-1198-2007 ยท GKZ Counter, Differential Cell

Abbott Laboratories

Abbott CELL-DYN 3000 Hematology Analyzer

DateAug 29, 2007 (initiated May 11, 2007)
StatusTerminated
ReasonThe product failed electrostatic discharge testing, which could result in product failure.
Root causeOther
ActionMandatory Technical Service Bulletins (TSB) are being issued to Field Service to upgrade all instruments in the field. All customers will be contacted by Field Service to schedule this mandatory upgrade. The upgrades are expected to take 12 months.
Quantity3,429 units, all models
DistributionWorldwide, including USA, Puerto Rico, Canada, Mexico, Argentina, Australia, Bahamas, Brazil, Chile, China, Colombia, Costa Rica, Dominican Republic, El Salvador, Ecuador, Germany, Guatemala, Hong Kong, Jamaica, Japan, New Zealand, Panama, Paraguay, Peru, Singapore, South Korea, Taiwan, Thailand, Ur
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K890491CELL-DYN 3000 MULTI-PARAMETER HEMATOLOGY ANALYER
Abbott Diagnostics, Apr 18, 1989
K955673CELL DYN 3000 MULTI-PARAMETER AUTOMATED HEMATOLOGY ANALYZER (MODIFICATION)
Abbott Laboratories, May 20, 1996
K980614ABBOTT CELL-DYN 3000 SERIES HEMATOLOGY SYSTEMS
Abbott Laboratories, Apr 24, 1998