Recall Z-1145-2007 ยท IYE Accelerator, Linear, Medical

Northwest Medical Physics Equipment Inc

Isoloc software Versions 6.5 and 4.5, Part Number NT-NW-425-520. Software for image guided radiation therapy.

DateAug 8, 2007 (initiated Jan 2, 2006)
StatusTerminated
ReasonIsoloc version 6.5 and 4.5 software issued a false warning message regarding Gantry angles when set to "Anatomic Landmark" setting.
Root causeSoftware design
ActionSoftware to fix the described problem was written and customers contacted by e-mail on 01/06/2006 that a software patch, titled 01022006 was available to their customers on the firm's internet website.
Quantity132
DistributionNationwide, Korea, Spain, and UK.