Recall Z-0079-2008 ยท NAY System, Surgical, Computer Controlled Instrument
Intuitive Surgical, Inc.
Intuitive Surgical, daVinci S Surgical System, Model Number: IS2000, A5.1 P1 and A5.0 P5 software level, Surgical Endoscopic Instrument Control System, Intuitive Surgical, Sunnyvale, CA 94086.
| Date | Feb 21, 2008 (initiated Mar 30, 2007) |
| Status | Terminated |
| Reason | System Lock-Up: Software anomalies could cause product failure during use, or on start-up. System transitions to a safe "soft-lock" state. |
| Root cause | Software design |
| Action | Intuitive Field Service Engineers installed a software upgrade beginning on 03/30/2007, and left a notification letter that the upgrade had been performed at each site. New software level is A5.1 P2. |
| Quantity | 159 |
| Distribution | Worldwide Distribution - USA, Belgium, China, France, India, Italy, Japan, Korea, Spain, Sweden, Switzerland, UK and Mexico. |