Recall Z-1163-2007 ยท ITI Wheelchair, Powered
Sunrise Medical Inc
Sunrise Medical Quickie Power Wheelchairs: Three models -Rhythm, Groove, and Zippy Z-Bop (with Quickie IQ Enhanced Displays)
| Date | Aug 14, 2007 (initiated Apr 23, 2007) |
| Status | Terminated |
| Reason | The wire harness connected to the display housing (Quickie IQ Enhanced Displays) may separate and cause sparks from a short circuit. |
| Root cause | Component design/selection |
| Action | Firm contacted consignees by phone and mailed May 1, 2007, Urgent Medical Device notification letters. |
| Quantity | 229 Enhanced Display units (affecting 220 wheelchairs) |
| Distribution | Product was distributed Worldwide. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K060531 | QUICKIE RHYTHM & GROOVE SERIES POWER WHEELCHAIRS W/QUICKIE DELPHI CONTROLS Sunrise Medical, Mar 17, 2006 |