Recall Z-1180-2007 ยท FPA Set, Administration, Intravascular

Carmel Pharma, Inc.

PhaSeal, Protector 21, Item No. 21, US Order No. 22020

DateAug 22, 2007 (initiated Jun 26, 2007)
StatusTerminated
ReasonRisk of Leakage: When the PhaSeal Protector and Injector are used with drug vials with thick rubber stoppers which have a concave upper surface and a flat bottom surface there is a potential for leakage. These rubber stoppers are made of a harder rubber formulation and may be more difficult to penetrate.
Root causeOther
ActionOn or about 6/15/2007, the firm sent advisory notice letters to the end users and the instructions for use have been updated by adding a precaution statement regarding the use of thick rubber stoppers. The firm is also updating the customer training program to include how to handle drug vials with thick concave rubber stoppers.
Quantity6.1 million for Z-1178-1181-2007 recalls
DistributionProduct was distributed throughout the US and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001368PROTECTOR 21, PROTECTOR 50, PROTECTOR 14, INJECTOR LUER LOCK, INFUSION ADAPTER