Recall Z-0717-2007 ยท KPR System, X-Ray, Stationary

Carestream Health, Inc.

Kodak Direct View DR3000 System, Stationary X-ray System, Catalog #s 655-1345, 655-1386.

DateJul 3, 2007 (initiated Mar 6, 2007)
StatusTerminated
ReasonThe x-ray systems do not indicate the field size when the system's detector is turned off and the collimator is in manual mode. This results in the systems being noncompliant with the field indication performance standards of section 1020.31(e)(2)& (3) of 21Code of Federal Regulations (CFR).
Root causeOther
ActionOn 3/06/07, Eastman Kodak notified CDRH of its proposed Corrective Action Plan (CAP). The CAP was approved by CDRH under the following conditions: (1) Kodak will visit each user facility, and apply dial indicator labels to the collimator knobs, apply a field size label to the collimator housing, and verify that each system complies with the field indication performance standard. (2) Kodak will prepare adequate instructions for the assembly/installation and testing of the dial indicators labels and field size labels. These instructions are to be furnished to all owners of the Kodak Direct View
QuantityDomestic: 7 units, Canada: 2 units
DistributionFL, NY, MO, MI, TX, and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060550KODAK DIRECTVIEW DR 3000 SYSTEM 64KW/80KW GENERATORS, MODELS 6551345, 6551386
Eastman Kodak Company, Mar 31, 2006