Recall Z-1127-2007 ยท GKP Instrument, Coagulation, Automated

Instrumentation Laboratory Co.

Instrumentation Laboratory ACL TOP Coagulation Analyzer (Base Model only). Blood analyzer. P/N 280000

DateAug 7, 2007 (initiated Jun 19, 2007)
StatusTerminated
ReasonACL TOP unit, when in emergency stop may release liquid during recovery sequence and contaminate sample tube.
Root causeOther
ActionInstrumentation Laboratory notiifed the distributor Beckman Coulter on 6/19/2007. Beckman Coulter notified their customers and provided a copy with the Instructions for Recovery Procedure on how to avoid contamination from probe leaks.
Quantity94
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063679ACL TOP
Instrumentation Laboratory CO, Jan 12, 2007