Recall Z-1148-2007 ยท MKT Hepatitis Viral B Dna Detection

Medtronic Emergency Response Systems, Inc.

biphasic LIFEPAK 500 automated external defibrillators (AEDs) with software version 4.4 or lower; Part Nos. 3011790-XXXXXX, D3100790-XXXXXX, and U3011790-XXXXXX . (XXXX indicates various non-contiguous numbers)

DateAug 4, 2007 (initiated Jun 13, 2007)
StatusTerminated
ReasonReduced Shock- the device may deliver 100 Joule (J) to a patient, which is less than the recommended factory minimum default setting of 200 J.
Root causeOther
ActionOn June 13, 2007 the recalling firm sent a letter dated June 11, 2007, titled URGENT - MEDICAL DEVICE CORRECTIVE ACTION. The letter advises of the situation, recommends to the consignee to set defibrillator to FIXED Sequence Protocol and to contact the firm for a FLEXIBLE Sequence Energy Protocol. The letter provides an instruction sheet and a response sheet to be faxed back to the company.
Quantity59,996 devices
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K955854LIFEPAK 500
Physio-Control Corp., Nov 4, 1996