ClearanceBrief
Recall Z-1112-2007 ยท FRN Pump, Infusion

Baxter Healthcare Corp.

Baxter Flo-Gard 6301 Dual Channel Volumetric Infusion Pump; Baxter Healthcare Corporation, Deerfield, IL 60015; Product code # 2M8064

DateJul 28, 2007 (initiated Jun 21, 2007)
StatusTerminated
ReasonFalsified Repair/Test and Inspection records: Pumps submitted for repair or service may not have been properly repaired or serviced before being returned to the customer.
Root causeOther
ActionBaxter telephoned the direct accounts starting on 6/21/07, informing them that functional verification testing may not have been performed on their Flo-Gard and/or Colleague infusion pumps before they were returned to them. Since Baxter has no evidence that this quality control testing was conducted on these pumps, the pumps may not function as designed. The accounts were requested to locate the affected pumps and take them out of service as soon as possible, and send them back to Baxter for the required functional verification testing. *** Baxter expanded the recall and telephoned the dire
Quantity144 pumps
DistributionNationwide: Alaska, Arizona, California, Colorado, Florida, Georgia, Hawaii, Idaho, Illinois, Iowa, Kansas, Kentucky, Louisiana, Maryland, Michigan, Mississippi, Montana, Nebraska, Nevada, New Jersey, New Mexico, New York, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, Tennessee
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K915523FLO-GARD(R) 6301
Baxter Healthcare Corp, Feb 28, 1992