Recall Z-1196-2007 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Sendx Medical Inc

ABL80 FLEX Analyzer, Model #393-839

DateAug 25, 2007 (initiated Mar 27, 2007)
StatusTerminated
ReasonThere have been reported experiences where waste fluids have sprayed out from the waste drain at the front of the ABL80 analyzer when excessive pressure builds within the waste line.
Root causeOther
ActionThe recall was initiated 3/27/2007. All distributors were notified via the Radiometer FAN (field action notice) system which provides a communication mechanism for these types of actions. The notice advises that there have been a few reported experiences where waste fluids have sprayed out from the waste drain at the front of the ABL80 analyzer which occurs when excessive pressure builds within the waste line. To prevent any spraying of fluids from the waste drain a check valve has been added to the waste tubing, between the waste drain and the waste pump which allows fluids to flow in one
Quantity334 units worldwide
DistributionClass II Recall - Worldwide Distribution --- USA including state of Ohio, and countries of Austria, Belgium, Denmark, Finland, France, Germany, Italy, Japan, The Netherlands, Norway, Poland, Spain, Sweden, Switzerland, United Kingdom, Australia, New Zealand, Canada, and Denmark.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.