Recall Z-1070-2007 ยท JHI Visual, Pregnancy Hcg, Prescription Use

Abbott Diagnostic International, Ltd.

Total B-hCG Speciment Diluent is composed of 1 bottle of Total B-hCG Specimen Diluent containing animal serum.

DateJul 12, 2007 (initiated Apr 10, 2007)
StatusTerminated
ReasonAn increase in background counts elevates the results of low level samples up to approximately 6mIU/mL. This is not enough to shift a negative result to positive on an undiluted sample. If low level samples are diluted then the background effect is magnified potentially yelding false positive results for pregnancy and or Dilution Inaccuracy.
Root causeOther
ActionThe 4/10/2007 Product Recall Letter was sent to all customers who received the affected lots.
Quantity1853
DistributionWorldwide, including USA, PR, USVI, CANADA, MEXICO, GUATEMALA, COSTA RICA, COLOMBIA, ECUADOR, PERU, CHILE, URUGUAY, ARGENTINA, BRAZIL, DOMINICAN REPUBLIC, EL SALVADOR, GERMANY, JAPAN, SINGAPORE, HONG KONG, TAIWAN, THAILAND, SOUTH KOREA, AUSTRALIA, NEW ZEALAND, HONDURAS, TRINIDAD & TOBAGO, GRENADA,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K935673AXSYM TOTAL B-HCG
Abbott Laboratories, Jun 21, 1994