Recall Z-1075-2007 ยท BZD Ventilator, Non-Continuous (Respirator)

Cardinal Health

AirLife nCPAP System Driver; a nasal continuous positive airway pressure device, which is part of a larger system that includes an nCPAP generator, fixation device and patient interface (prongs and.or mask); Cardinal Health, McGaw Park, IL 60085; cat. 006900

DateJul 12, 2007 (initiated Jun 1, 2007)
StatusTerminated
ReasonUnder isolated conditions, the device may experience issues with O2 Calibration, which may result in either a nuisance alarm or a delivered FiO2 of up to 5% below the specified tolerance.
Root causeOther
ActionCardinal Health sent the accounts recall letters dated 6/1/07 via UPS next day delivery, which listed the catalog number and affected serial numbers of the CPAP Driver, and informed them that under certain isolated conditions, the CPAP Driver may experience issues with O2 Calibration, which may result in either a nuisance alarm or a delivered FiO2 of up to 5% below the specified tolerance. Customers have experienced this under the following conditions: 1) air inlet pressure is greater than O2 pressure by 5 to 10 psi and 2) FiO2 settings less than 45%. Separate recall letters were sent to th
Quantity48 units
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051226AIRLIFE NASAL CONTINUOUS POSITIVE AIRWAY PRESSURE (NCPAP) SYTEM DRIVER
Cardinalhealth, Nov 17, 2005