Recall Z-1074-2007 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems Inc

Varian Medical Systems Clinac Medical Linear Accelerator. Clinac 600 C &C/D, Clinac 2100 C, C/D & EX and Clinac 2300 C &C/D

DateJul 12, 2007 (initiated May 2, 2007)
StatusTerminated
ReasonThe wedge tray may slip out of the Clinac interface mount during gantry rotation.
Root causeProcess control
ActionThe firm sent out a Product Notification Letter on 5/2/07 to customers who have ordered this guide block through a Varian Service Center. A Varian Field Engineer will visually inspect guide blocks at-site that have been installed from Service stock for all affected customers, and replace if necessary.
Quantity209 units
DistributionProduct was distributed and manufactured between 1999 and 2006. Distributed worldwide, including USA, Ecuador, Columbia, Taiwan, Austrailia, Austria, Brazil, Canada, Chile, China, Finland, France, Germany, India, Italy, Japan, New Zealand, Poland, Portugal, Serbia and Montenegro, Singapore, Spain,