Recall Z-1165-2007 ยท FMK Single Use Only Blood Lancet With An Integral Sharps Injury Prevention Feature

Lifescan Inc

One Touch SureSoft Lancing Device in carton. Part Number 021-140-01, aka OT SureSoft Hospital, Gentle 200

DateAug 16, 2007 (initiated Mar 30, 2007)
StatusTerminated
ReasonLancet tip may not fully retract and cause needle stick and exposure to used lancets.
Root causeNonconforming Material/Component
ActionThe firm has conducted a health hazard evaluation, and has initiated a corrective action plan. The firm initated recall on March 30, 2007, and notifications were sent to hospital end users known to Lifescan and its direct accounts. In addition, a customer service script has been developed to address any questions from incoming users. The firm has requested direct accounts to perform subrecalls on Lifescan's behalf to notify customers of the recall as needed.
Quantity632 cartons (200 units per carton)
DistributionProduct was distributed nationwide to distributors and wholesale dealers.