Recall Z-1193-2007 ยท FLE Sterilizer, Steam

Steris Corp

Amsco Century Medium Steam Sterilizer (26" x 37.5") (660 x 950mm).

DateAug 23, 2007 (initiated Mar 22, 2007)
StatusTerminated
ReasonThe cables inside the pump conduit box may be exposed (not covered with wire nut and/or electrical tape) that could cause a potential electrical shock hazard.
Root causeEmployee error
ActionBy telephone 3/22/2007, customers were instructed to cease using the product and a service technician would be dispatched immediately to repair the unit.
Quantity78
DistributionWorldwide, including USA, Denmark, Belgium, Greece, India, and Russia.There are no Canadian accounts.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010865AMSCO CENTURY MEDIUM STERILIZER, MODEL CENTURY
STERIS Corporation, May 31, 2001