Recall Z-1042-2007 ยท NBW System, Test, Blood Glucose, Over The Counter

Lifescan Inc

OneTouch Ultra Test strip, a component of the following OneTouch Ultra Glucose monitoring systems: a)OneTouch Ultra, b) OneTouch InDuo, c) OneTouch Ultra2, d) OneTouch UltraMini, e) OneTouch UltraSmart. 25 strips per vial. Product comes in packs of 100 and 50. (Part numbers: 020-245-07, 020-244-07, 020-963-01)

DateJul 13, 2007 (initiated Mar 22, 2007)
StatusTerminated
ReasonDamaged Vials/Inaccurate test results: a manufacturing equipment problem on one of the packaging lines had the potential to puncture vials, causing a single hole in the side of the vial below the rim. This result of this defect may produce inaccurately high or low blood glucose test results or error message readings.
Root causeOther
ActionMarch 2007 via letters to consumers and consignees and web postings. The firm has also set up verbal communication plans for customers who contact Lifescan by phone.
Quantity~33 million vials (25 strips per vial in the US, ~19 million worldwide.
DistributionWorldwide; USA, Europe, Middle East, Africa, Latin America and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062195MODIFICATION TO ONE TOUCH ULTRA BLOOD GLUCOSE MONITORING SYSTEM
Lifescan, Inc., Aug 16, 2006