Recall Z-1157-2007 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Datascope Corporation

The ClearGlide EVH small with Scissors. The kit includes three Precision Bipolar Devices, an Optical Vessel Dissector, a Small Ultra Retractor, a Vessel Dissector, and the ENDOPATH 5mm curved scissors. Datascope Cardiac Assist Sterile Manufactured for: Datascope Cardiac Assist 15 Law Drive Fairfield, NJ 07004

DateAug 9, 2007 (initiated May 7, 2007)
StatusTerminated
ReasonDatascope has experienced a higher than usual number of reports involving the Precision Bipolar Device. Reports include the locking or sticking of the buttons of the device which operate the jaws, particularly when a side load is placed on the shaft of the device.
Root causeOther
ActionLetters were sent to all hospitals, (contact: Hospital Administrators) the week of May 7, 2007 (certified mail, Return Receipt requested) with Urgent Medical Device Field Correction. The letter requests that all potential users of the devices at the institution be mdae aware of the situation. The letter instructs clinicians what to do if they experience problems with locking buttons of the Precision Device and also if the device is clamped closed on a vein branch and cannot be opened. Institutions are also requested to sign and return a postal card acknowledging receipt of the Field Correct n
Quantity549
DistributionWorldwide, including USA, Belgium, the Netherlands, and Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003587CLEARGLIDE PRECISION BIPOLAR DEVICE
ETHICON, Inc., Dec 4, 2000