Recall Z-1073-2007 ยท FFK Lithotriptor, Electro-Hydraulic

Gyrus ACMI Corporation

Gyrus CyberWand Ultrasonic Lithotripter System includes: CW-USLS CyberWand USL System CW-USLT CyberWand USL Transducer CW-USLRBP CyberWand USL Renal/Bladder Probe Set CW-USLTT CyberWand USL Transducer Sterilization Tray

DateAug 9, 2007 (initiated May 23, 2007)
StatusTerminated
ReasonSterilization Instructions not adequate for sterilizing.
Root causeOther
ActionGyrus has informed their sales force to visit each customer that has a CyberWand system(s) and hand deliver a letter, dated May 23, 2007, that explains the situation and requests that the customer cease using the unit and segregate it to prevent accidental use. The letter also explains that the firm is collecting any sterilization trays with printed sterilization instructions and replacing them. The rep will assist the customer in complying. The customer will also be required to complete a reply form to track compliance.
Quantity53 units
DistributionNationwide and Canada.
FDA recordOpen on fda.gov