Recall Z-1009-2007 ยท FPO Stretcher, Wheeled

Stryker Medical Div. of Stryker Corporation

Stryker Medical -Wheeled Electric Stretcher, Model 1550 (with and without scales)

DateSep 21, 2007 (initiated May 7, 2007)
StatusTerminated
ReasonThe internal power cord may become damaged due to wear and, if this occurs, a shock hazard exists.
Root causeOther
ActionConsignees were notified of the problem via letter dated 5/7/07 instructing them to immediately unplug the affected stretchers with scales, affix a warning label enclosed with the letter to the power cord of stretchers with scales, and that they will be contacted so that the firm can schedule a visit to upgrade all affected stretchers. A second notification letter was issued on August 7, 2007 to cover all 1550 stretchers with and without scales.
Quantity1,048 (1009 in the U.S. plus 39 international)
DistributionWorldwide: USA, Latin America, Mexico and Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.