Recall Z-1008-2007 ยท HBB Motor, Drill, Pneumatic

The Anspach Effort, Inc.

Anspach Single Use, Sterile Bone Cutting Burrs

DateJul 11, 2007 (initiated May 18, 2007)
StatusTerminated
ReasonThe white film portion of the sterile pouch may become brittle prior to the expiry date compromising the sterility of the product.
Root causeNonconforming Material/Component
ActionThe firm issued a recall letter on May 18, 2007, to its consignee, advising them of the recall. Consignees were asked to screen their inventory for the recalled devices and complete an accompanying product replacement form. The firm will replace all returned product within 30 days.
Quantity1,962,123
DistributionWorld wide-Distributed throughout the United States and to Puerto Rico, Canada, Mexico, Japan, Lithuania, Cyprus, Austrailia, England, Sweden, Bahamas, Korea, Germany, Switzerland, Malaysia, Iceland, Portugal, Singapore, Poland, Hungary, Thailand, Austria, Israel, Belgium, Egypt, South Africa, Nethe
FDA recordOpen on fda.gov