Recall Z-0991-2007 ยท BTL Ventilator, Emergency, Powered (Resuscitator)

Smiths Medical PM, Inc.

Pneupac VR1 Standard and VR1 Responder is a hand held portable, time cycled, gas powered, flow generator ventilatory resuscitators that are intended for emergency resuscitation by medical personnel, paramedics and ambulance technicians inside and outside hospital and for ventilation by medical personnel inside and outside hospital in emergency situations and for intra- and inter- hospital transpor

DateJul 4, 2007 (initiated May 16, 2007)
StatusTerminated
ReasonSmiths Medical is conducting a recall of 93 Pneupac VR1 Emergency & Transport Ventilators following the identification of a problem which may exist with the demand valve that could cause the ventilator to stop cycling & to deliver continuous flow output to the patient.
Root causeOther
ActionDistriblutors and end users were contacted by phone (May 10 - May 24, 2007) and end users were additionally sent a Safety Action Bulletin May 21, 2007. The Bulletin was used as a script.
Quantity93
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051322PNEUPAC VRI STANDARD, MODEL VRISTDNUS, PNEUPAC VRI RESPONDER, MODEL VRIRESNUS
Smiths Medical International , Ltd., Aug 5, 2005