ClearanceBrief
Recall Z-0985-2007 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Toshiba American Med Sys Inc

Toshiba Aplio 80 Digital Ultrasound System, SSA-770A with version 6 software

DateJun 27, 2007 (initiated Feb 2, 2007)
StatusTerminated
ReasonWhen using the LV measurement function and making numerous measurements the value displayed on the image screen may on occasion not update correctly. The values that are stored in MDA (measurement display area) are correct and no issue has been reported with these values.
Root causeOther
ActionNotification letters were sent to customers on February 2, 2007. Customers were informed of the safety issue regarding the possibility of incorrect measurement values being displayed on the active image screen while performing multiple measurements and instructed to verify actual results against the values stored in the MDA (measurement display area). Customers were also informed that a Toshiba Customer Service Engineer will contact them to schedule a software upgrade when software becomes available. If customers have any questions they were instructed to contact their Toshiba Customer Service
Quantity260
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041499SSA-770A, APLIO VERSION 5.5
Toshibamedical Systems Corporation, Jun 10, 2004