ClearanceBrief
Recall Z-0984-2007 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips Medical Systems North America Co. Phillips

Panorama 1.0T MRI

DateJun 27, 2007 (initiated Apr 30, 2007)
StatusTerminated
ReasonPotential for gaseous helium to improperly vent, resulting in rupture of the magnet vessel and venting of helium into the MRI site.
Root causeOther
ActionOn 4/30/07 Philips sent an Urgent Device Correction letter dated April 25, 2007 via certified mail to customers. The letter advises customers of the issue and that that a representative will visit the sites, inspect and correct the problem if indicated.
Quantity56 units
DistributionUnits were distributed throughout the US to 1,078 hospitals and medical centers.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041602PANORAMA 1.0T
Philips Medical Systems North America, Inc., Jul 26, 2004