Recall Z-1007-2007 ยท LZG Pump, Infusion, Insulin

Medtronic MiniMed

Medtronic MiniMed Paradigm Insulin Infusion Pumps for the following Paradigm models: MMT-511, MMT-512, MMT-712, MMT-515, MMT-715, MMT-522 and MMT-722

DateJul 7, 2007 (initiated Apr 27, 2007)
StatusTerminated
ReasonExposure to Magnetic Resonance Imaging (MRI) has resulted in damage to the component that monitors and controls movement of the motor in the MiniMed Paradigm insulin infusion pump. Although there were alarms as a result of the damage, some users cleared these alarms and continued using the pump. Under such conditions, the pump will significantly over-deliver; potentially causing severe hypoglycemi
Root causeOther
ActionFirm mailed Medical Device Safety letters on April 24, 2007, to healthcare professionals, existing pump users and MRI facilities reiterating the existing warning in the pump User Guide to avoid exposing the pump to MRI (or similar high strength electromagnetic fields) and strengthening previous warning by specifically mentioning the potential for over-delivery and severe hypoglycemia. Firm is also including an insert with this information with any Paradigm infusion pumps shipped to new customers.
Quantity334,000 Paradigm insulin pumps (approx)
DistributionWorldwide: USA, US Virgin Islands, Antigua, Australia, Austria, Bahamas, Bahrain, Bermuda, Brazil, Canada, Chile, China, Columbia, Czech Republic, Denmark, Dominican Republic, Egypt, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Kuwait, Lebanon, Malaysia, Mexico,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001829MINIMED PARADIGM INSULIN PUMP, MODEL 511
Minimed, Inc., Aug 18, 2000
K030531MEDTRONIC MINIMED PARADIGM MODEL 512 INSULIN PUMP AND THE BD PARADIGM LINK GLUCOSE METER
Minimed, Inc., Jun 17, 2003
K031390MEDTRONIC MINIMED PARADIGM INSULIN PUMP, MODEL MMT-712
Minimed, Inc., Jul 23, 2003
K040676MEDTRONIC MINIMED PARADIGM INSULIN PUMP, MODELS MMT-515 AND MMT-715
Medtronic Minimed, May 21, 2004
K053177MEDTRONIC MINIMED PARADIGM INSULIN PUMP, MODEL MMT-712E
Medtronic Minimed, Jan 31, 2006