ClearanceBrief
Recall Z-1037-2007 ยท LOM Test, Hepatitis B (B Antigen, B Surface Antigen, Be Antigen)

Abbott Laboratories

AxSYM CORE 2.0 Reagent Pack; list 08B88; 100 test pack; in-vitro diagnostic; Abbott Laboratories, Abbott Park, IL 60064

DateJul 11, 2007 (initiated Apr 23, 2007)
StatusTerminated
ReasonFailure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
Root causeOther
ActionAbbott initiated a field correction recall by letter dated 4/23/07 to all customers receiving the affected reagents. The accounts were informed of the failure of the AxSYM to open (actuate) some reagent packs with the affected lots. The accounts were requested to inspect each reagent pack for broken or improperly sealed caps; manually open each bottle to ensure the flipper bar raises the reagent cap correctly; close each flipper bar prior to loading the reagent pack on the instrument; and discard any reagent packs that do not open properly. If the user observes a probe crash, prior to testing
Quantity70 packs
DistributionWorldwide, including USA, Puerto Rico, Argentina, Australia, Brazil, Canada, Cayman Islands, Chile, Colombia, Ecuador, Germany, Hong Kong, Mexico, New Zealand, Singapore, South Korea, Taiwan, Thailand and Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.