Recall Z-0812-2007 ยท GKP Instrument, Coagulation, Automated

Diagnostica Stago, Inc.

STA Compact and STA Compact CT For in vitro Diagnostic Use Only In vitro diagnostic analyzer Diagnostica Stago France

DateMay 16, 2007 (initiated Jan 18, 2007)
StatusTerminated
ReasonResult of complaints of field failures of the silicon bridge rectifier, resulting in smoke billowing out of the unit and burnt circuit boards, wires and connectors.
Root causeOther
ActionA recall notification has not been sent. The firm did issue a technical bulletin to their representatives.--Update: based on FDA input the firm issued a Recall Letter to all users dated May 25, 2007
Quantity693 units
DistributionProduct was distributed to hospitals and laboratories nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K961579STA COMPACT AUTOMATED MULTI-PARAMETRIC ANALYZER
American Bioproducts Co., Jul 10, 1996