Instrumentarium Dental, Inc.

Cranex Tome panoramic x-ray unit with Spiral Tomography for cross-sectional imaging. manufactured by SOREDEX, PO Box 250, 00031 Helsinki, Finland.

DateMay 18, 2007 (initiated Apr 28, 2006)
StatusTerminated
ReasonThere may be a material defect in the up/down movement axle of the unit. This may cause the upper part of the unit to suddenly come down.
Root causeOther
ActionA recall letter, dated 04/28/2006, was sent to affected customers . The letter describes the issue, and requests that no patient operation is permitted until the unit is checked by authorized service technician. If the service technician noted wear of the screw, the assembly was replaced.
Quantity9
DistributionWorldwide, including USA, Canada, Bermuda, and Colombia.