ClearanceBrief
Recall Z-0781-2007 ยท FWF Camera, Television, Endoscopic, Without Audio

Smith & Nephew, Inc Endoscopy Division

SOPRO Camera Control Units, NTSC, SV420P; REF (Catalog) #72200505 v1.1; Made in France; Manufactured for Smith & Nephew, Inc., Endoscopy, Andover, MA 01810.

DateMay 5, 2007 (initiated Jan 19, 2007)
StatusTerminated
ReasonSurgical camera control units were not designed and manufactured to pass the IEC 60101 testing for electrical fast transients (EFT)
Root causeOther
ActionRecalling firm sent Recall notification letters to customers (Jan 2007) and distribution centers requesting device return.
Quantity77 units.
DistributionForeign Distribution Only-No U. S. A. distribution International distribution to: Argentina, Austria, Australia, Belgium, Brazil, China, Columbia, Denmark, Dubai, France, Germany, Greece, Italy, India, Japan, Korea, Malaysia, Netherlands, Norway, Portugal, South Africa, Spain, Sweden, Turkey and U
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.