Recall Z-0933-2007 ยท MCX Catheter, Coronary, Atherectomy

St Jude Medical CRMD

St. Jude Medical-CPS Direct SL- slittable CRT lead delivery tool. -Model Numbers 410110, 410111, 410112, 410113, 410114, 410115, 410116, 410120,410121. 410122, 410123, 410124, 410125 and 410126.

DateJun 14, 2007 (initiated Mar 20, 2007)
StatusTerminated
ReasonDevice Separation- product may separate at the mapping collar transition joint located at the proximal (light blue) marker band end of the tool during the implant procedure.
Root causeOther
ActionBy Letter dated March 20, 2007, customers were requested to returned affected product immediately to St. Jude Medical. St. Judes Sales Representative will be assisting customers in retrieving the subject inventory from their shelves. The firm instructed customers that if they should require further information regarding this issue, to contact the firm's Technical Service department at (800) 722-3774 or customers can contact local St. Jude Medical representative.
Quantity14,000
DistributionWorldwide-including states of AZ, CA DC, FL GA, IN, KS, LA, MA, MD, ME, MI, MN, MS, NC, NH, NY, OH, PA, RI, TN, VA, and WI and countries of Austria, Belgium, Denmark, Finland, Netherlands, Sweden, France, Germany, Italy, UK, Spain, Portugal, Poland, Switzerland, Australia, New Zealand, Thailand, Hon
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K053277CPS DIRECT SL OUTER CATHETER
St Jude Medical, Feb 24, 2006