Recall Z-0978-2007 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems Inc

LINAC Scapel Treatment Planning System, incorporating FastPlan software versions 4.0, 5.0, 5.0.1, 5.1, 5.1.1.

DateJun 28, 2007 (initiated Mar 9, 2007)
StatusTerminated
ReasonA software anomaly in a medical device used for patient treatment could result in accidental or unintended exposure radiation exposure. The tilt rotation compensation algorithm does not adequately compensate for the tilt rotation.
Root causeSoftware design
ActionThe recalling firm notified medical device end users by letter dated 03/09/07 and advised of the software anomaly and provided details for user corrective action pending software upgrade by the manufacturer.
Quantity80 devices
DistributionThe medical devices were distributed to medical centers nationwide and worldwide.